Selection of Common Materials

Insulation Material Group
(IEC 60112)
CTI
(IEC 60112)
PLC/PTI
(UL746)
Typical materials
I >600 0 PTFE; PE; PA; PFA; FEP; Makrofol® (PC, E168120); Formex® (PP, E121855); Kapton® 300FN929
II 400….599V 1 Polyester (PET)
IIIa 250….399 V 2
IIIa 175….249 V 3 PPS, PEN
IIIb 100….174 V 4 Polyimide; PEI; PSU; PEEK
< 100V 5

Insulation coordination according to IEC 60664, like many UL standards, uses the CTI value, which is the most common. Relevant standards include, for example, IEC 60587 and IEC 60112, as well as UL Standard 749.

Classification into insulating material groups according to IEC 60601:

CTI value
Insulating material group
600 ≤ CTI I
400 ≤ CTI < 600 II
175 ≤ CTI < 400 IIIa
100 ≤ CTI < 175 IIIb

    In addition to the basic standard DIN IEC 60664 (Insulation Coordination), there is a draft application guide (design examples and insulation tests) for Part 2 (DIN IEC 60664-2-1).

    When determining the PTI (Proof Tracking Index) and the CTI (Comparative Tracking Index), the IEC 60112 standard proceeds as follows:

    In the first step, a conductive liquid is dripped between two platinum electrodes resting flat on the surface of the insulating material (up to 50 drops every 30 seconds). This simulates condensation and surface contamination.

    The voltage is gradually increased until either the current exceeds 0.5 A or ignition and erosion occur. The highest voltage at which the test specimen does not fail is verified at five points. This voltage is the PTI value (the test rating).

    In the second step, the determined voltage is reduced by the PTI value and 25 V. Using this voltage, the material is tested at five different locations with 50 or 100 drops every 30 seconds. If the test specimen passes the test, this voltage then becomes the determined comparative number (CTI value).

    It is important to note that higher frequencies are not covered by the standard, yet they have a significant influence on the formation of a creep path. Electrochemical migration is also not taken into account due to the relatively short test duration (50 × 30 seconds = 25 minutes).

    In many cases, the CTI value in “volts” and the unitless PTI test number are used interchangeably (e.g., CTI = 1 actually corresponds to CTI = 400…599 V or Insulation Material Group II).